COVID-19 & Influenza A+B Antigen Combo Test Kit (Colloidal Gold)

Cassette (1 test/kit,2 tests/kit,20 tests/kit,25tests/kit,40 tests/kit,50 tests/kit)

All products use the Cassette format OEM/ODM

Width options are consistent (3.0mm or 4.0mm)

All specimens are compatible with Nasal Swab Test

The COVID-19 & Influenza A+B Antigen Combo Test Kit (Colloidal Gold) is a rapid visual immunoassay for the qualitative detection of nucleocapsid protein antigen (N antigen) from SARS-CoV-2 and influenza A+B viral antigen (nucleoprotein) in human throat swab and nasal swab samples, only for in vitro diagnostic use.

1.  Specimen extraction

(1). Find the extraction tube and carefully tear off the sealed foil film on the extraction tube. Do not let the extraction buffer flow out, and put it on the work station(or insert the tube into the hold on the kit box.).

(2). Insert the swab sample into the extraction tube which contains extraction buffer. Roll the swab at least 6 times while pressing the head against the bottom and side of the extraction tube.  

(3). Leave the swab in the extraction tube for one minute.  

(4). Remove the swab while squeezing the sides of the tube to extract the liquid from the swab. The extracted solution will be used as test sample.

(5). Insert the dropper tip into the extraction tube tightly.

COVID-19 & Influenza A+B Antigen Combo Test Kit (Colloidal Gold)

(The picture is for references only, please refer to the material object.)

2. Detection operations:

  1. Open a pouch containing a test cassette. Place the test cassette on a dry, horizontal work surface.
  2. Add 2 drops (about 60μl) of sample solution extract to the sample well respectively of the test cassette.
  3. Observe the results showed within 10-15 minutes, and the results showed after 15 minutes have no clinical significance.
  4. Negative: One colored line appears in the control region (C). No apparent colored line appears in the test region.

    Positive: Two or three distinct colored lines appear. One colored line should be in the control region (C) and the other colored line(s) should be in the test region. 

    Invalid: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test with a new test cassette. If the problem persists, discontinue using the test kit immediately and contact your local distributor.